<div style="display:inline;"> <img height="1" width="1" style="border-style:none;" alt="" src="//googleads.g.doubleclick.net/pagead/viewthroughconversion/972768237/?value=0&amp;guid=ON&amp;script=0"/> </div>
 

Celtac
Fully-automatic hematology analyzer with 33 parameters

  • Up to 90 samples per hour

  • 33 parameters

  • Micro sampling capability

  • Continuous loading of samples via rack fed system

  • Onscreen shortcut keys allows easy operation.

  • up to 7 racks of 10 tubes

  • STAT/manual sample analysis

  • Laser scatter + flow cytometry technology

  • Built-in rocking mixing

  • Smart ColoRac Match system

  • Integrated validation station with touch screen

  • Reagent and controls management with barcode


  • Parameters

  • Reagents

  • Throughput

  • Precision

  • (Reproducibility)

  • Complete

  • QC programs

  • Options

  • Dimensions and Weight

  • 33: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV,
    RDW-SD, PLT, PCT, MPV, PDW, P-LCR,
    NE, NE%, LY, LY%, MO, MO%, EO, EO%, BA, BA%,
    P-LCC*, Mentzer Index*, RDWI*,
    IG%*, IG#*, Band%*, Band#*, Seg%*, Seg#*
    (*) RUO parameters (Research Use Only)

  • 5: 2 hemolyzing reagents,
    1 diluent, 2 detergents

  • CBC : 25 μL
    CBC+DIFF : 40 μL
    Pre-dilution mode : 20 μL

  • Up to 90 samples /hour

  • WBC : 2.0% or less (WBC : 4.00 x 103/μL or more)
    RBC : 1.5% or less (RBC : 4.00 × 106/μL or more)
    HGB : 1.5% or less
    HCT : 1.5% or less
    MCV : 1.0% or less
    PLT : 4.0% or less
    (PLT : 100 x 103/μL or more)
    NE% : 5.0% or less
    (NE% : 30.0% or more AND WBC :
    4.00 x 103/μL or more)
    LY% : 5.0% or less
    (LY% : 15.0% or more AND WBC :
    4.00 x 103/μL or more)
    MO% : 12.0% or less
    (MO% : 5.0% or more AND WBC :
    4.00 x 103/μL or more)
    EO% : 20.0% or less OR within ±1.0 EO%
    (WBC : 4.00 × 103/μL or more)
    BA% : 30.0% or less OR
    within ±1.0 BA%
    (WBC : 4.00 x 103/μL or more)
    NE : 8.0% or less (NE :
    1.20 x 103/μL or more)
    LY : 8.0% or less (LY :
    0.60 x 103/μL or more)
    MO : 20.0% or less
    (MO : 0.20 x 103/μL or more)
    EO : 25.0% or less OR within
    ±0.10 x 103/μL
    (WBC : 4.00 x 103/μL or more)
    BA : 30.0% or less OR
    within 0.10 x 103/μL (WBC : 4.00 x 103/μL or more)
    (Specifications above applies
    to normal mode)

  • L & J / Westgard Multi rule /
    X bar B(X bar Batch) management

  • Software kit MEK-9100, QS-023W
    Extra sample racks, T411A
    Waste sensor, JW-910W

  • 675 W x 589 D x 576 H mm, 66 kg